Xpress Flex® Proper Techniques for Testing
STEPS | FIGURES |
---|---|
Pour your sample water into the center chamber on the test vial (~80 mL). |
|
Carefully measure prep reagents for Iron Prep and Phosphate Prep, and add to respective chambers. • 0.5 mL of phosphate reagent to chamber 4. |
|
Use gloves and protective eyewear when handling Iron and Phosphate prep reagents. Immediately clean up any spilled reagents. |
|
Carefully use the syringe to measure sample water from center chamber of test vial and add to chambers 1 and 4. • Add 5 mL of sample water to both chambers. |
|
Cap center chamber of test vial and swirl to mix. |
|
Dip timing is important. Pay attention to the sounds and animation as instructed. Remove promptly. Avoid flinging water. |
|
When dipping strips, be sure numbers on clip and testing vial line up. |
|
Avoid handling test strips near pads when loading into Xpress Flex. |
|
Follow all program prompts. |
|
Never dip into the same sample twice. |
|
Keep lids of the test strip bottles closed when not in use. |
|
If Xpress Flex shuttle starts accumulating water, use a dry cloth to remove. |