the most trusted name in water testing

K-0431

Reagent Pack, Buret, Alkalinity (P/M&P/T), Phenolphthalein/Methyl Orange/Blended Indicator, 1mL=5mg

UPC Barcode:
840036010136

$40.10

Analyte System Method/Chemistry Standard/Equivalance or Description Comparator Cell
Alkalinity, P/M & P/T Buret titration Phenolphthalein/methyl orange/blended indicator 1 mL = 5 mg P/M or P/T alkalinity as CaCO₃ NA NA
Test Parameter Description
High halogen level may change indicator reaction from green/red to blue/yellow; to prevent, add thiosulfate prior to testing. Filming amines, quats, and biguanides may interfere and cause the endpoint to look purple; to prevent, add 10 drops of R-0884 prior to testing. Alkalinity

REAGENT SHELF LIFE

All reagents have a shelf life, whether they are liquids, powders, crystals, tablets, or test-strip pads. If kept dry, powders and crystals are very stable; acids are also long lived. Date of manufacture is not the controlling factor when it comes to shelf life—storage conditions are more important. As with all perishables, reagents are sensitive to environmental influences and will last longer under controlled conditions.

To this end, we recommend:

  • Storing reagents at a consistent temperature in the range if 36°–85°F (2°–29°C); extreme temperature fluctuation, say from a refrigerator to a hot car trunk, causes reagents to deteriorate.
  • Keeping them out of prolonged direct sunlight. (Note: their brown plastic bottles help protect very light-sensitive reagents.)
  • Segregating reagents from containers of treatment chemicals.
  • Replacing caps immediately and tightening them carefully so that exposure to air and humidity is limited.
  • Avoiding switching bottle caps, placing bottle caps on soiled surfaces, repouring reagents into contaminated containers, or touching test strip pads.

Taylor formulates its reagents to remain effective for at least one year, with only very few exceptions (molybdenum indicator in liquid form is one; after four months old it should be tested against a standard periodically). As a general precaution, replace all reagents more than one year old, or at the beginning of a new testing season.