Xpress Flex® Troubleshooting


Troubleshooting: 

ISSUE POSSIBLE CAUSE  RESULT  CORRECTIVE ACTION

Lot Code for test strip(s) was not updated with new bottle

Test results may be inconsistent/unusual 

Update test strip lot code in Xpress Flex sureTREAT program.

Sample could have low Total Alkalinity

pH is inconsistent or different than another test

Increase alkalinity of pool/spa and return for a retest.

CYA pad on Flex 3 did not reach sample during dipping

CYA reading will be inaccurate

Retest with fresh sample and proper technique to ensure pad reaches sample.

No prep reagent added to chamber 1

Iron test gives repeated <0.15 ppm result

Retest with fresh sample and use 1 scoop of Iron Prep Reagent (R-9103).

Reinstallation of XFConnect is required

Flex Strip lot codes do not save 

Please contact Taylor Technical Support at 877-837-8548 for assistance.

 


Xpress Flex Error Codes: 

ERROR CODES MESSAGE

WrongSlot: Flex 1 detected but NOT in position 1

WrongSlot: Flex 3 detected but NOT in position 3

Retest: PO4 Prep missing. Add Prep when re-testing PO4

Invalid: Pad color not recognized

Note: If Invalid alerts appear, verify that the lot code on your bottle matches the lot code selected in sureTREAT®

Invalid: Pad color not recognized

Note: If Invalid alerts appear, verify that the lot code on your bottle matches the lot code selected in sureTREAT®

Alert: Total Alkalinity too low (<20 ppm) to measure pH or CYA

Alert: Br is too high

 


 Xpress Flex Lighted Color Codes:  

CODE COLOR SEQUENCE